Model:Flame AA System-A3F, Graphite AA System-A3G, Combined Flame and Graphite Furnace AA System-A3FG | |
Spectrometer System | |
Wavelength range | 190-900 nm |
Light source | Hollow cathode lamp (HCL), Deuterium Arc lamp (D2) |
Modulation | Square wave pulse |
Modulation frequency | 100 hz self reversal (SR) background, 400 hz Deuterium Arc (D2) background |
Monochromator | |
Grating | 1800 grooves/mm diffraction grating |
Blazed wavelength | 250 nm |
Focus | 300 nm |
Bandwidth | 0.1, 0.2, 0.5, 1.0, 2.0, 5.0 nm |
Scan mode | Automatic |
Photometric type | Single beam |
Wavelength accuracy | ±0.15 nm |
Wavelength resolution | 0.2 nm±0.02 nm |
Wavelength reproducibility | <0.05 nm |
Spectral bandwidth | 0.5, 1.0, 2.0, 5.0 nm |
Baseline stability | 0.005 A/30 min |
Stray light | ≤0.12 %T (220 nm NaI, 340 nm NaNO2) |
Background correction | Deuterium Arc (D2) 1.0 Abs, Self Reversal (SR) 3.0 Abs |
Flame Analysis | |
Flame types | Air/Acetylene, Nitrous Oxide/Acetylene, Air/Propane (LPG) |
Sensitivity (Cu) | 2 μg/ml Absorption >0.280 Abs |
Characteristic concentration | Cu<0.02 μg/ml, Ba<0.15 μg/ml (N2O/Acetylene) |
Detection limit | Cu<0.004 μg/ml |
Repeatability | Cu<0.7 % (Air/Acetylene Flame) Ba<1.0 % (Nitrous Oxide/Acetylene Flame) |
Burner heads | Titanium alloy |
Nebulizer | High efficiency glass, Acid proof available as an option |
Pre-mix chamber | Corrosion resistant |
Atomizer selection | Manual (A3F) Automatic changeover (A3FG) |
Safety features | Burner identification, flame sensor, gas leak sensor, low gas pressure sensor, drain trap sensor, power loss protection |
Graphite Furnace Analysis | |
Graphite head | Transversally heated |
Temperature range | Ambient-2650 ℃ |
Heating program | Up to 10 steps. Drying, Ashing, Atomization, Cleaning |
Feedback | Voltage and optical temperature control feedback |
Sensitivity (Cu) | 50 ng/m Absorption>0.40 Abs |
Detection limit | Cd<0.004 ng/ml |
Repeatability | Cu<2.0 %, Cd<2.0 % |
Graphite tubes | Pyrolytically coated with platform |
Sample size | Up to 20μl |
Graphite cooling | Water circulator available |
Safety features | Burner Identification, Flame Sensor, Gas leak Sensor, Low Gas Pressure Sensor, Drain Trap Sensor, Power Loss Protection, Circulation Water (graphite), Over Temperature Sensor (graphite) |
Data Processing | |
Operating program | AA Win Pro software |
Analytical methods | Flame AA, Flame AE, Graphite Furnace, Hydride Generation |
Readout Mode | Continuous, manual, peak height, peak area |
calibrations | Multi-standard calibration, standard addition, interpolation |
Data storage | Analytical results, instrument and measurement parameters, signal profile, calibration curve |
Power Requirements | |
Main unit | 220 VAC 50/60 Hz 200 W |
Graphite power supply | 380VAC 50/60Hz, instantaneous power 5 kW current rating 7 kW |
Dimensions | |
Main unit | 44 x 21 x 21 in 165 lb |
Graphite power supply | 20 x 21 x 21 in 154 lb |